Endoscopic Imaging Using a Multimode Optical Fibre Calibrated with Multiple Internal References

  1. 1.
    SYSNO ASEP0552078
    Document TypeJ - Journal Article
    R&D Document TypeJournal Article
    Subsidiary JČlánek ve WOS
    TitleEndoscopic Imaging Using a Multimode Optical Fibre Calibrated with Multiple Internal References
    Author(s) Jákl, Petr (UPT-D) RID, ORCID, SAI
    Šiler, Martin (UPT-D) RID, ORCID, SAI
    Ježek, Jan (UPT-D) RID, ORCID, SAI
    Cifuentes, Angel S. (UPT-D) RID, SAI, ORCID
    Traegaardh, Johanna (UPT-D) RID, ORCID, SAI
    Zemánek, Pavel (UPT-D) RID, SAI, ORCID
    Čižmár, Tomáš (UPT-D) RID, ORCID, SAI
    Article number37
    Source TitlePhotonics. - : MDPI
    Roč. 9, č. 1 (2022)
    Number of pages10 s.
    Publication formOnline - E
    Languageeng - English
    CountryCH - Switzerland
    Keywordsfibre endoscopy ; spatial light modulator ; transmission matrix
    Subject RIVBH - Optics, Masers, Lasers
    OECD categoryOptics (including laser optics and quantum optics)
    R&D ProjectsEF15_003/0000476 GA MŠMT - Ministry of Education, Youth and Sports (MEYS)
    LO1212 GA MŠMT - Ministry of Education, Youth and Sports (MEYS)
    ED0017/01/01 GA MŠMT - Ministry of Education, Youth and Sports (MEYS)
    Method of publishingOpen access
    Institutional supportUPT-D - RVO:68081731
    UT WOS000759106500001
    EID SCOPUS85123507712
    DOI10.3390/photonics9010037
    AnnotationThe interferometric acquisition of the transmission matrix (TM) of a multimode optical fibre (MMF), which is at the heart of multimode fibre-based endoscopic imaging methods, requires using a reference beam. Attempts to use an internal reference, that is to provide the reference in a common pathway geometry through the MMF itself, lead to a speckled reference intensity and consequential occurrence of “blind spots” - locations where insufficient optical power in the reference wave inflicts strong measurement errors. Here we show that combining a relatively small number of TMs, which are measured using different internal references, facilitates a complete elimination of blind spots, and thereby a significant enhancement of the imaging quality.
    WorkplaceInstitute of Scientific Instruments
    ContactMartina Šillerová, sillerova@ISIBrno.Cz, Tel.: 541 514 178
    Year of Publishing2023
    Electronic addresshttps://www.mdpi.com/2304-6732/9/1/37