Number of the records: 1  

First simultaneous measurements of Reynolds stress with ball-pen and Langmuir probes

  1. 1.
    SYSNO ASEP0476139
    Document TypeJ - Journal Article
    R&D Document TypeJournal Article
    Subsidiary JČlánek ve WOS
    TitleFirst simultaneous measurements of Reynolds stress with ball-pen and Langmuir probes
    Author(s) Grover, Ondřej (UFP-V) ORCID
    Adámek, Jiří (UFP-V) RID, ORCID
    Seidl, Jakub (UFP-V) RID
    Devitre, A. (CR)
    Šos, Miroslav (UFP-V) ORCID
    Vondráček, Petr (UFP-V) RID, ORCID
    Bílková, Petra (UFP-V) RID
    Hron, Martin (UFP-V) RID, ORCID
    Article number063501
    Source TitleReview of Scientific Instruments. - : AIP Publishing - ISSN 0034-6748
    Roč. 88, č. 6 (2017)
    Number of pages8 s.
    Publication formPrint - P
    Languageeng - English
    CountryUS - United States
    KeywordsCOMPASS tokamak ; Electric field effects ; Electric fields ; Langmuir probes ; Magnetoplasma ; Ball-pen probes
    Subject RIVBL - Plasma and Gas Discharge Physics
    OECD category2.11 Other engineering and technologies
    R&D ProjectsGA15-10723S GA ČR - Czech Science Foundation (CSF)
    GA16-25074S GA ČR - Czech Science Foundation (CSF)
    LM2015045 GA MŠMT - Ministry of Education, Youth and Sports (MEYS)
    Institutional supportUFP-V - RVO:61389021
    UT WOS000404641300015
    EID SCOPUS85020257420
    DOI10.1063/1.4984240
    AnnotationA new multi-pin probe head was installed and tested at the COMPASS tokamak. The probe head consists of several ball-pen and Langmuir probes in similar geometric configurations, which enables fast, simultaneous measurements of the radial and poloidal electric field components from differences of plasma or floating potentials using appropriately positioned ball-pen or Langmuir probes, respectively. The fast measurements of the radial electric field are compared with other methods
    of measuring selected frequency components of the radial electric field. The radial profiles of the Reynolds stress calculated from correlations of electric field fluctuations obtained by either probe type are compared along with their spectral composition. Lower Reynolds stress values are found for Langmuir probes in comparison to ball-pen probes due to negative contributions from higher frequency fluctuations possibly associated with electron temperature fluctuations.
    WorkplaceInstitute of Plasma Physics
    ContactVladimíra Kebza, kebza@ipp.cas.cz, Tel.: 266 052 975
    Year of Publishing2018