Number of the records: 1  

Direct, rapid quantitative analyses of BVOCs using SIFT-MS and PTR-MS obviating sample collection

  1. 1.
    SYSNO ASEP0361441
    Document TypeJ - Journal Article
    R&D Document TypeJournal Article
    Subsidiary JČlánek ve WOS
    TitleDirect, rapid quantitative analyses of BVOCs using SIFT-MS and PTR-MS obviating sample collection
    Author(s) Smith, D. (GB)
    Španěl, Patrik (UFCH-W) RID, ORCID
    Source TitleTrAC-Trends in Analytical Chemistry. - : Elsevier - ISSN 0165-9936
    Roč. 30, č. 7 (2011), s. 945-959
    Number of pages15 s.
    Languageeng - English
    CountryGB - United Kingdom
    KeywordsSIFT-MS ; PTR-MS ; Absolute quantification
    Subject RIVCF - Physical ; Theoretical Chemistry
    CEZAV0Z40400503 - UFCH-W (2005-2011)
    UT WOS000294318200013
    DOI10.1016/j.trac.2011.05.001
    AnnotationThe purpose of this short review is to describe the origins and the principles of operation of selected-ion flow-tube mass spectrometry (SIFT-MS) and proton-transfer-reaction mass spectrometry (PTR-MS), and their application to the analysis of biogenic volatile organic compounds (BVOCs) in ambient air, the humid air (headspace) above biological samples, and other samples. We briefly review the ion chemistry that underpins these analytical methods, which allows accurate analyses. We pay attention to the inherently uncomplicated sampling methodologies that allow on-line, real-time analyses, obviating sample collection into bags or onto traps, which can compromise samples.
    WorkplaceJ. Heyrovsky Institute of Physical Chemistry
    ContactMichaela Knapová, michaela.knapova@jh-inst.cas.cz, Tel.: 266 053 196
    Year of Publishing2012