Number of the records: 1
Intercomparison of radiation protection protection devices in a high-energy stray neutron field. Part III: Instrument response
- 1.
SYSNO ASEP 0331468 Document Type J - Journal Article R&D Document Type Journal Article Subsidiary J Článek ve WOS Title Intercomparison of radiation protection protection devices in a high-energy stray neutron field. Part III: Instrument response Title Srovnání zařízení radiační ochrany v rozptýleném neutronovém poli o vysoké energii. Část III: Odezva přístroje Author(s) Silari, M. (CH)
Agosteo, S. (IT)
Beck, P. (AT)
Bedogni, R. (IT)
Cale, E. (FR)
Caresana, M. (IT)
Domingo, C. (ES)
Donadille, L. (FR)
Dubourg, N. (FR)
Esposito, A. (IT)
Fehrenbacher, G. (DE)
Fernández, F. (ES)
Ferrarini, M. (IT)
Fiechtner, A. (CH)
Fuchs, A. (CH)
García, M. J. (ES)
Golnik, N. (PL)
Gutermuth, F. (DE)
Khurana, S. (DE)
Klages, T. (DE)
Latocha, M. (AT)
Mares, V. (DE)
Mayer, S. (CH)
Radon, T. (DE)
Reithmeier, H. (DE)
Rollet, S. (AT)
Roos, H. (DE)
Rühm, W. (DE)
Sandri, S. (IT)
Schardt, D. (DE)
Simmer, G. (DE)
Spurný, František (UJF-V)
Trompier, F. (FR)
Villa-Grasa, C. (FR)
Weitzenegger, E. (DE)
Wiegel, B. (DE)
Wielunski, M. (DE)
Wissmann, F. (DE)
Zechner, A. (AT)
Zielczyński, M. (PL)Source Title Radiation Measurements. - : Elsevier - ISSN 1350-4487
Roč. 44, 7-8 (2009), s. 673-691Number of pages 19 s. Language eng - English Country GB - United Kingdom Keywords radiation protection devices ; radiation field ; detectors ; dosemeters Subject RIV BG - Nuclear, Atomic and Molecular Physics, Colliders R&D Projects IAA100480902 GA AV ČR - Academy of Sciences of the Czech Republic (AV ČR) CEZ AV0Z10480505 - UJF-V (2005-2011) UT WOS 000271688400004 DOI 10.1016/j.radmeas.2009.05.005 Annotation Work Package 6 of CONRAD dealt with ‘‘complex mixed radiation fields at workplaces’’ and in this context it organised a benchmark exercise, which included both measurements and calculations, in a stray radiation field at a high-energy particle accelerator at GSI, Germany. The aim was to intercompare the response of several types of active detectors and passive dosemeters in a well-characterised workplace field. This paper focuses on the intercomparison of the response of the dosemeters in terms of ambient dose equivalent. The paper describes in detail the detectors employed in the experiment, followed by a discussion of the results. A comparison is also made with the H*(10) values predicted by the Monte Carlo simulations and those measured by the BSS systems. Workplace Nuclear Physics Institute Contact Markéta Sommerová, sommerova@ujf.cas.cz, Tel.: 266 173 228 Year of Publishing 2010
Number of the records: 1