SYSNO ASEP 0324963 Document Type J - Journal Article R&D Document Type Journal Article Subsidiary J Ostatní články Title Compensating the Effect of Static Magnetic Field in MR Measurement of Diffusion Title Kompenzace vlivu statického magnetického pole v MR měřeních difúze Author(s) Bartušek, Karel (UPT-D) RID, ORCID, SAI
Gescheidtová, E. (CZ)Number of authors 2 Source Title PIERS Online - ISSN 1931-7360
Roč. 5, č. 1 (2009), s. 86-90Number of pages 5 s. Language eng - English Country US - United States Keywords MRI ; static magnetic field ; diffusion Subject RIV JA - Electronics ; Optoelectronics, Electrical Engineering R&D Projects GA102/07/0389 GA ČR - Czech Science Foundation (CSF) CEZ AV0Z20650511 - UPT-D (2005-2011) DOI 10.2529/PIERS080820070632 Annotation The paper describes a modified PFGSE MR method of measuring diffusion coefficients in heterogeneous materials. The principle of the proposed method consists in measuring the decrease in spin echo when applying diffusion gradients of different polarities. The detected images must be processed using special methods. The method greatly reduces the effect of static magnetic field gradient on the precision of measuring diffusion coefficients. Workplace Institute of Scientific Instruments Contact Martina Šillerová, sillerova@ISIBrno.Cz, Tel.: 541 514 178 Year of Publishing 2009