Number of the records: 1  

Hyphenated ultra high-performance liquid chromatography–Nano Quantity

  1. 1.
    0327080 - MBÚ 2010 RIV NL eng J - Journal Article
    Olšovská, Jana - Kameník, Zdeněk - Cajthaml, Tomáš
    Hyphenated ultra high-performance liquid chromatography–Nano Quantity.
    Journal of Chromatography A. Roč. 1216, č. 30 (2009), s. 5774-5778. ISSN 0021-9673. E-ISSN 1873-3778
    R&D Projects: GA MŠMT 1M06011
    Institutional research plan: CEZ:AV0Z50200510
    Keywords : macrolides * uv detection * ultra high-perfonmance liquid chromatography
    Subject RIV: EE - Microbiology, Virology
    Impact factor: 4.101, year: 2009

    A novel universal aerosol-based detector Nano Quantity Analyte Detector – NQADTM, connected with an ultra-performance liquid chromatography system is described. The detector was employed for detection of selected antibiotic compounds – macrolides (oleandomycin, erythromycin, troleandomycin, clarithromycin and roxithromycin) that are hard to detect using classical UV detectors due to the lack of chromophores. The determined lowest detection limits under isocratic conditions for these compounds ranged from 3.0 to 5.4 μg/mL. The suitability of the detector connected with ultra high-performance liquid chromatography in the gradient mode was tested on a more complex mixture containing 12 antibiotics. The detector exhibited full compatibility under both the elution modes when UHPLC separations were achieved in relatively short run times
    Permanent Link: http://hdl.handle.net/11104/0173980

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.