Number of the records: 1  

Capillary electrophoresis with contactless conductivity detection for rapid determination of meldonium in urine

  1. 1.
    SYSNO ASEP0467992
    Document TypeC - Proceedings Paper (int. conf.)
    R&D Document TypeConference Paper
    TitleCapillary electrophoresis with contactless conductivity detection for rapid determination of meldonium in urine
    Author(s) Šlampová, Andrea (UIACH-O) RID
    Kubáň, Pavel (UIACH-O) RID, ORCID
    Number of authors2
    Source TitleCECE 2016. 13th International Interdisciplinary Meeting on Bioanalysis. - Brno : Ústav analytické chemie AV ČR, v. v. i., 2016 / Foret František ; Křenková Jana ; Drobníková Iveta ; Klepárník Karel - ISBN 978-80-904959-4-4
    Pagess. 234-236
    Number of pages3 s.
    Publication formOnline - E
    ActionCECE 2016. International Interdisciplinary Meeting on Bioanalysis /13./
    Event date17.10.2016 - 19.10.2016
    VEvent locationBrno
    CountryCZ - Czech Republic
    Event typeWRD
    Languageeng - English
    CountryCZ - Czech Republic
    Keywordsmeldonium ; capillary electrophoresis-capacitively coupled contactless conductivity detection ; urine
    Subject RIVCB - Analytical Chemistry, Separation
    R&D ProjectsGA16-09135S GA ČR - Czech Science Foundation (CSF)
    Institutional supportUIACH-O - RVO:68081715
    AnnotationA simple, rapid and cheap capillary electrophoresis-capacitively coupled contactless
    conductivity detection (CE-C4D) method for determination of meldonium (MEL) in
    urine samples was developed. Sample pretreatment was minimized to dilution of urine
    samples with DI water and MEL was determined by CE-C4D in background electrolyte
    consisting of 2 M acetic acid (pH 2.3) in ng mL−1 to mg mL−1 concentration range.
    Sample-to-sample analysis time was less than 4 min, limit of detection and
    quantification was 0.015 and 0.05 μg mL−1, respectively, and the method showed
    excellent linearity (r2 ≥ 0.9998), recovery (97.6–99.9%) and intraday as well as interday
    repeatability (RSD ≤ 4.7%).