Instument Intercomparison in the High-energy Mixed Field at the CERN-EU Reference Field (CERF) Facility
- 1.0435788 - ÚJF 2015 RIV GB eng J - Journal Article
Caresana, M. - Helmecke, M. - Kubančák, Ján - Manessi, G. P. - Ott, K. - Scherpelz, R. - Silari, M.
Instument Intercomparison in the High-energy Mixed Field at the CERN-EU Reference Field (CERF) Facility.
Radiation Protection Dosimetry. Roč. 161, 1-4 (2014), s. 67-72. ISSN 0144-8420. E-ISSN 1742-3406
Institutional support: RVO:61389005
Keywords : aircrew dosimetry * CERF * CERN
Subject RIV: BG - Nuclear, Atomic and Molecular Physics, Colliders
Impact factor: 0.913, year: 2014
This paper discusses an intercomparison campaign performed in the mixed radiation field at the CERN-EU (CERF) reference field facility. Various instruments were employed: conventional and extended-range rem counters including a novel instrument called LUPIN, a bubble detector using an active counting system (ABC 1260) and two tissue-equivalent proportional counters (TEPCs). The results show that the extended range instruments agree well within their uncertainties and within 1 sigma with the H*(10) FLUKA value. The conventional rem counters are in good agreement within their uncertainties and underestimate H*(10) as measured by the extended range instruments and as predicted by FLUKA. The TEPCs slightly overestimate the FLUKA value but they are anyhow consistent with it when taking the comparatively large total uncertainties into account, and indicate that the non-neutron part of the stray field accounts for similar to 30 % of the total H*(10).