Number of the records: 1  

Calibration of germanium temperature sensors of helium cryostat for experimental study of natural turbulent convection

  1. 1.
    0352300 - ÚPT 2011 RIV CZ eng C - Conference Paper (international conference)
    Urban, Pavel - Hanzelka, Pavel - Musilová, Věra
    Calibration of germanium temperature sensors of helium cryostat for experimental study of natural turbulent convection.
    The 11th Cryogenics 2010 - IIR International Conference. Praha: ICARIS Ltd, 2010, s. 183-185. ISBN 978-2-913149-77-9.
    [Cryogenics 2010 /11./. Bratislava (SK), 26.04.2010-29.04.2010]
    R&D Projects: GA MŠMT ED0017/01/01; GA AV ČR KJB200650902
    Institutional research plan: CEZ:AV0Z20650511
    Keywords : comparative calibration * calibration device * temperature sensors calibration * natural convection * Rayleigh number
    Subject RIV: BJ - Thermodynamics

    Permanent Link: http://hdl.handle.net/11104/0191836
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.