Number of the records: 1  

Quantitation of trace analytes in capillary zone electrophoresis with UV-absorbance detection: A critical study of approaches based on peak height

  1. 1.
    0162416 - UIACH-O 20020068 RIV DE eng J - Journal Article
    Křivánková, Ludmila - Gebauer, Petr - Pantůčková, Pavla - Boček, Petr
    Quantitation of trace analytes in capillary zone electrophoresis with UV-absorbance detection: A critical study of approaches based on peak height.
    Electrophoresis. Roč. 23, č. 12 (2002), s. 1833-1843. ISSN 0173-0835. E-ISSN 1522-2683
    R&D Projects: GA ČR GA203/99/0044; GA ČR GA203/01/0401; GA ČR GA203/02/0023; GA AV ČR IAA4031703; GA AV ČR IAA4031103
    Institutional research plan: CEZ:AV0Z4031919
    Keywords : capillary zone electrophoresis * quantitative analysis * peak height evaluation
    Subject RIV: CB - Analytical Chemistry, Separation
    Impact factor: 4.325, year: 2002

    It was shown that the shape and slope of calibration curves measured for analytes in matrices containing stacking or non-stacking constituents clearly show whether the conditions selected for the analysis give correct results and higher attainable sensitivity. Rules for successful quantitation in samples with complex matrices were defined.