Number of the records: 1
Quantitation of trace analytes in capillary zone electrophoresis with UV-absorbance detection: A critical study of approaches based on peak height
- 1.0162416 - UIACH-O 20020068 RIV DE eng J - Journal Article
Křivánková, Ludmila - Gebauer, Petr - Pantůčková, Pavla - Boček, Petr
Quantitation of trace analytes in capillary zone electrophoresis with UV-absorbance detection: A critical study of approaches based on peak height.
Electrophoresis. Roč. 23, č. 12 (2002), s. 1833-1843. ISSN 0173-0835. E-ISSN 1522-2683
R&D Projects: GA ČR GA203/99/0044; GA ČR GA203/01/0401; GA ČR GA203/02/0023; GA AV ČR IAA4031703; GA AV ČR IAA4031103
Institutional research plan: CEZ:AV0Z4031919
Keywords : capillary zone electrophoresis * quantitative analysis * peak height evaluation
Subject RIV: CB - Analytical Chemistry, Separation
Impact factor: 4.325, year: 2002
It was shown that the shape and slope of calibration curves measured for analytes in matrices containing stacking or non-stacking constituents clearly show whether the conditions selected for the analysis give correct results and higher attainable sensitivity. Rules for successful quantitation in samples with complex matrices were defined.
Permanent Link: http://hdl.handle.net/11104/0059731
Number of the records: 1