Determination of the interdetector volume by s-detection in size exclusion chromatography of polymers with on-line multiangle light-scattering detection
- 1.0105699 - UMCH-V 20043269 RIV NL eng J - Journal Article
Netopilík, Miloš - Podzimek, Š. - Kratochvíl, Pavel
Determination of the interdetector volume by s-detection in size exclusion chromatography of polymers with on-line multiangle light-scattering detection.
[Stanovení mezidetektorového objemu s-detekcí v rozměrově vylučovací chromatografii polymerů s on-line víceúhlovou detekcí rozptylu světla.]
Journal of Chromatography A. Roč. 1045, 1-2 (2004), s. 37-41. ISSN 0021-9673. E-ISSN 1873-3778
R&D Projects: GA AV ČR(CZ) KSK4050111
Keywords : instrument parameters * gel permeation chromatography * multiangle light scattering detection
Subject RIV: CD - Macromolecular Chemistry
Impact factor: 3.359, year: 2004
The slope of the local calibration found by the w-detection is very sensitive to the interdetector-volume error, whereas the slope obtained by the s-detection, using data from the light-scattering detector only, does not require knowledge of the interdetector volume. The correct IDV value yields a local calibration found by the w-detection with the same slope as the local calibration determined by the s-detection. In this way, a sufficiently accurate value of IDV can be found.
Směrnice lokální kalibrace nalezené w-detekcí je velmi citlivá k chybě v mezidetektorovém objemu, zatímco směrnice získaná s-detekcí, pouze za využití dat rozptylového detektoru, nevyžaduje znalost mezidetektorvého objemu. Správný mezidetektorvý objem je takový, který poskytuje lokální se stejnou směrnicí jako lokální kalibrace stanovená s-detekcí. Tímto způsobem je možno získat dostatečně přesnou hodnotu mezidetektorového objemu.