Number of the records: 1
Sampling procedure and a ratio-indicator study of mercury determination in whole blood by using an AMA 254 atomic absorption spectrometer
- 1.0102291 - UIACH-O 20043066 RIV DE eng J - Journal Article
Spěváčková, V. - Korunová, Vlasta - Čejchanová, M. - Vobecký, Miloslav
Sampling procedure and a ratio-indicator study of mercury determination in whole blood by using an AMA 254 atomic absorption spectrometer.
[Postup přípravy vzorků a radio-indikátorová studie stanovení rtuti v celé krvi atomovým absorpčním spektrometrem AMA 254.]
Analytical and Bioanalytical Chemistry. Roč. 380, č. 2 (2004), s. 346-350. ISSN 1618-2642. E-ISSN 1618-2650
R&D Projects: GA ČR GA203/01/0453
Institutional research plan: CEZ:AV0Z4031919
Keywords : mercury * ratio-indicator * blood samples stability
Subject RIV: CB - Analytical Chemistry, Separation
Impact factor: 2.098, year: 2004
A sampling procedure appropriate for the mercury determination on whole blood was tested. The inactive controls and rat blood with metabolised 197Hg ratio-indicator were used. A key factor in the correct determination of mercury in whole blood was found to be the vessel material. The polypropylene vessels were found to be suitable among the vessel types tested. The function of the instrument was checked by ratio-indicator 197Hg
Byl testován vhodný způsob uchování vzorků plné krve pro stanovení rtuti. Byla použita krysí krev obsahující metabolizovaný radioaktivní-indikátor 197Hg a referenční vzorek lidské krve. Klíčovým faktorem správného stanovení rtuti v celé krvi byl materiál nádobek, ve kterých byla krev uchovávána. Jako nejvhodnější materiál byl potvrzen polypropylen. Přístoj AMA 254 byl pomocí radio-indikátoru testován na možné ztráty rtuti během stanovení
Permanent Link: http://hdl.handle.net/11104/0009646