Characterisation of neutron fields around high-energy x-ray radiotherapy machines
- 1.
SYSNO ASEP 0100850 Document Type J - Journal Article R&D Document Type Journal Article Subsidiary J Ostatní články Title Characterisation of neutron fields around high-energy x-ray radiotherapy machines Title Charakteristiky neutronových polí v okolí radioterapeutických zařízení s vysokoenergetickým zářením X Author(s) Králík, M. (CZ)
Turek, Karel (UJF-V) ORCID, SAISource Title Radiation Protection Dosimetry. - : Oxford University Press - ISSN 0144-8420
Roč. 110, 1-4 (2004), s. 503-507Number of pages 5 s. Language eng - English Country GB - United Kingdom Keywords radiotherapy machines ; neutron fields ; high-energy Subject RIV DN - Health Impact of the Environment Quality CEZ AV0Z1048901 - UJF-V Annotation Photoneutron spectra around the treatment bed of a Varian Clinac 2100C machine were measured using a Bonner sphere spectrometer. To overcome problems with pulse pile-up and detection of non-neutron-induced events, the active detector of thermal neutrons normally used at the centre of the spheres was replaced by a sandwich of four CR-39 track detectors interleaved with 10B radiators. Track densities measured for the CR-39 detectors in Bonner spheres were used for the unfolding of neutron spectra. Neutron fluence and ambient dose equivalent for the whole energy range and partial energy intervals were derived from the neutron spectra Workplace Nuclear Physics Institute Contact Markéta Sommerová, sommerova@ujf.cas.cz, Tel.: 266 173 228 Year of Publishing 2005