Number of the records: 1
Magnetic susceptibility measurement using 2D magnetic resonance imaging
- 1.0366067 - ÚPT 2012 RIV GB eng J - Journal Article
Marcon, P. - Bartušek, Karel - Burdkova, M. - Dokoupil, Zdeněk
Magnetic susceptibility measurement using 2D magnetic resonance imaging.
Measurement Science and Technology. Roč. 22, č. 10 (2011), 105702:1-8. ISSN 0957-0233. E-ISSN 1361-6501
R&D Projects: GA ČR GAP102/11/0318; GA MŠMT ED0017/01/01
Institutional research plan: CEZ:AV0Z20650511
Keywords : magnetic flux density * magnetic susceptibility * MRI * MR signal * reaction field
Subject RIV: JA - Electronics ; Optoelectronics, Electrical Engineering
Impact factor: 1.494, year: 2011
The authors describe a nuclear magnetic resonance (NMR) method for measuring the magnetic susceptibility of bar-shaped samples that have an arbitrary cross-section and do not produce an MR signal. The method is based on the measurement of the 2D map of the reaction field in the vicinity of a non-ferromagnetic sample and on the calculation of magnetic susceptibility from a known reaction field. The verification of the technique was realized via modelling the measured configuration in the ANSYS program by means of the finite element method and through an experimental measurement of MR-compatible and MR-incompatible materials carried out on a tomograph. A great advantage of the proposed susceptibility evaluation method consists in the use of only standard commercially used devices without the need of any special sequences.
Permanent Link: http://hdl.handle.net/11104/0201158
Number of the records: 1