SYSNO ASEP 0162227 Document Type J - Journal Article R&D Document Type Journal Article Subsidiary J Ostatní články Title Fate of free selenium atoms in externally heated quartz tube atomizers for hydride generation atomic absorption spectometry and their reatomization at tube ends studied by means of the determination of longitudinal free atom distribution Author(s) Matoušek, Tomáš (UIACH-O) RID, ORCID
Dědina, Jiří (UIACH-O) RID, ORCIDSource Title Spectrochimica Acta Part B: Atomic Spectroscopy. - : Elsevier - ISSN 0584-8547
Roč. 55, - (2000), s. 545-557Number of pages 3 s. Language eng - English Country NL - Netherlands Subject RIV CB - Analytical Chemistry, Separation R&D Projects GA203/98/0754 GA ČR - Czech Science Foundation (CSF) KSK2055603 GA AV ČR - Academy of Sciences of the Czech Republic (AV ČR) Annotation Longitudinal distribution of Se free atoms in heated quartz tube hydride atomizers was studied under analytical conditions. The free atom decay rate is greatly enhanced at higher analyte concentrations, which is a source of calibration curvature. The presence of zones where once reacted atoms are reatomized at the tube ends was directly proven: the sensitivity is improved and calibration bending is attenuated. Recommendations concerning the quartz tube design and dimensions are given. Workplace Institute of Analytical Chemistry Contact Iveta Drobníková, drobnikova@iach.cz, Tel.: 532 290 234 Year of Publishing 2001