Number of the records: 1  

An Improved Apparatus for Vapor–liquid Equilibria Measurement in Systems Containing Polymer at Higher Temperatures.

  1. 1.
    SYSNO ASEP0474799
    Document TypeC - Proceedings Paper (int. conf.)
    R&D Document TypeThe record was not marked in the RIV
    TitleAn Improved Apparatus for Vapor–liquid Equilibria Measurement in Systems Containing Polymer at Higher Temperatures.
    Author(s) Pavlíček, Jan (UCHP-M) RID, SAI
    Bogdanić, Grozdana (UCHP-M) RID, ORCID, SAI
    Wichterle, Ivan (UCHP-M) RID, ORCID, SAI
    Source TitleBook of Abstracts. - -, 2017 - ISBN N.
    Pagess. 236
    Number of pages1 s.
    Publication formOnline - E
    ActionEuropean Symposium on Applied Thermodynamics - ESAT 2017 /29./
    Event date18.05.2017 - 21.05.2017
    VEvent locationBucharest
    CountryRO - Romania
    Event typeWRD
    Languageeng - English
    CountryRO - Romania
    Keywordsvapor-liquid equilibrium ; polymer ; solvent
    Subject RIVCF - Physical ; Theoretical Chemistry
    OECD categoryPhysical chemistry
    R&D ProjectsGA15-19542S GA ČR - Czech Science Foundation (CSF)
    Institutional supportUCHP-M - RVO:67985858
    AnnotationThe apparatus for total pressure measurement in polymer + solvent systems has been designed recently. In the presented work, the range of functioning of this apparatus was extended to allow investigations at higher temperatures. A direct capacitance pressure sensor (Baratron) was replaced by a pressure transducer, which may be heated up to 200 oC. In addition, the solutions with higher concentrations of polymer can be studied in contrast to the ebulliometric method, which is limited because of problems with physical boiling of viscous mixtures.
    WorkplaceInstitute of Chemical Process Fundamentals
    ContactEva Jirsová, jirsova@icpf.cas.cz, Tel.: 220 390 227
    Year of Publishing2018
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.