Number of the records: 1
Effects of selected operational parameters on efficacy and selectivity of electromembrane extraction. Chlorophenols as model analytes
- 1.
SYSNO ASEP 0431195 Document Type J - Journal Article R&D Document Type Journal Article Subsidiary J Článek ve WOS Title Effects of selected operational parameters on efficacy and selectivity of electromembrane extraction. Chlorophenols as model analytes Author(s) Šlampová, Andrea (UIACH-O) RID
Kubáň, Pavel (UIACH-O) RID, ORCID
Boček, Petr (UIACH-O) RIDNumber of authors 3 Source Title Electrophoresis. - : Wiley - ISSN 0173-0835
Roč. 35, č. 17 (2014), s. 2429-2437Number of pages 9 s. Publication form Print - P Language eng - English Country DE - Germany Keywords electromembrane extraction ; chlorophenols ; extraction selectivity Subject RIV CB - Analytical Chemistry, Separation R&D Projects GA13-05762S GA ČR - Czech Science Foundation (CSF) Institutional support UIACH-O - RVO:68081715 UT WOS 000341701800002 DOI 10.1002/elps.201400096 Annotation A comprehensive quantitative study, based also on measurements of electric charge passed through the EME system, was carried out, which demonstrates that apart from the pH value, also the nature of counter-ions in donor and acceptor solution plays a significant role in the electrically induced transfer of charged analytes across supported liquid membranes. Workplace Institute of Analytical Chemistry Contact Iveta Drobníková, drobnikova@iach.cz, Tel.: 532 290 234 Year of Publishing 2015
Number of the records: 1