Number of the records: 1  

First-Data-Point problem and the baseline distortion in Fourier-transformNMR spectroscopy with simultaneous sampling

  1. 1.
    SYSNO ASEP0205517
    Document TypeJ - Journal Article
    R&D Document TypeJournal Article
    Subsidiary JOstatní články
    TitleFirst-Data-Point problem and the baseline distortion in Fourier-transformNMR spectroscopy with simultaneous sampling
    Author(s) Starčuk jr., Zenon (UPT-D) RID, ORCID, SAI
    Bartušek, Karel (UPT-D) RID, ORCID, SAI
    Starčuk, Zenon (UPT-D) RID
    Source TitleJournal of Magnetic Resonance. A - ISSN 1064-1858
    Roč. 108, - (1994), s. 177 - 188
    Number of pages12 s.
    Languageeng - English
    CountryUS - United States
    Keywordsbaseline distortion ; NMR
    Subject RIVJA - Electronics ; Optoelectronics, Electrical Engineering
    CEZAV0Z2065902 - UPT-D
    AnnotationThe baseline distortions often arising in spectra after linear phase correction have often been attributed to an improper value of the first acquired data point of the free-induction decay. However, correction of this value by multiplying it by a certain factor depending on the timing of the excitation sequence has not proved to be reliable unless some very specific conditions are met. In this article, the origin of these distortions is thoroughly examined in both the frequency and the time domain. It is shown that the problem can be explained generally as an improper resampling of the original signal as a result of the linear phase correction. A way of avoiding these artifacts is proposed. However, it is shown that generally theproper correction must affect the whole signal in a complex way and that suitable timing of the experiment is highly desirable.
    WorkplaceInstitute of Scientific Instruments
    ContactMartina Šillerová, sillerova@ISIBrno.Cz, Tel.: 541 514 178
    Year of Publishing2003

Number of the records: 1  