Number of the records: 1  

Radiation hardness tests of SiPMs for CBM PSD readout - in lab and at NA61@CERN beamtime

  1. 1.
    0489237 - ÚJF 2019 DE eng V - Research Report
    Mikhaylov, Vasily - Kugler, Andrej - Kushpil, Vasilij - Morozov, S. - Svoboda, Ondřej
    Radiation hardness tests of SiPMs for CBM PSD readout - in lab and at NA61@CERN beamtime.
    Darmstadt: GSI helmholtzzentrum für Schwerionenforschung GmbH, 2017. 486 s. GSI Scientific Report, GSI Scientific Report 2016. ISSN 0174-0814
    Institutional support: RVO:61389005
    Keywords : FAIR * GSI * Hadronic physics
    OECD category: Nuclear physics

    During the SiPM irradiation tests, Hamamatsu S12572-010P was chosen as the best available candidate for the PSD readout. Dark current increase with the absorbed fluence raises the requirement for the corresponding highcurrent power supply up to 1 mA per SiPM (channel). Gradual degradation of the SiPM signal to noise ratio with the absorbed fluence poses the requirement for the regular PSD module recalibration during the experiment operation.
    PSD energy resolution R_E studied at NA61 beamtime dropped only slightly - by 4.6 % for the SiPMs irradiated by total fluence of 4E11 n/cm2.