Number of the records: 1
Compensating the Effect of Static Magnetic Field in MR Measurement of Diffusion
- 1.0324963 - ÚPT 2009 RIV US eng J - Journal Article
Bartušek, Karel - Gescheidtová, E.
Compensating the Effect of Static Magnetic Field in MR Measurement of Diffusion.
[Kompenzace vlivu statického magnetického pole v MR měřeních difúze.]
PIERS Online. Roč. 5, č. 1 (2009), s. 86-90. ISSN 1931-7360
R&D Projects: GA ČR(CZ) GA102/07/0389
Institutional research plan: CEZ:AV0Z20650511
Keywords : MRI * static magnetic field * diffusion
Subject RIV: JA - Electronics ; Optoelectronics, Electrical Engineering
The paper describes a modified PFGSE MR method of measuring diffusion coefficients in heterogeneous materials. The principle of the proposed method consists in measuring the decrease in spin echo when applying diffusion gradients of different polarities. The detected images must be processed using special methods. The method greatly reduces the effect of static magnetic field gradient on the precision of measuring diffusion coefficients.
Článek popisuje modifikovanou MR metodu Pulsed Field Gradient Spin Echo pro měření difúzních koeficientů v heterogenních materiálech. Principem navržené metody je měření poklesu spinového echa při aplikaci difúzních gradientů různých polarit a navrženého zpracování měřených obrazů. Metoda výrazně snižuje vliv statického gradientu magnetického pole a křížového člene na přesnost měření difúzních koeficientů.
Permanent Link: http://hdl.handle.net/11104/0172536
Number of the records: 1