Number of the records: 1  

Gaseous Kr-83m generator for KATRIN

  1. 1.
    0489281 - ÚJF 2019 RIV GB eng J - Journal Article
    Sentkerestiová, Jana - Dragoun, Otokar - Lebeda, Ondřej - Ryšavý, Miloš - Sturm, M. - Vénos, Drahoslav
    Gaseous Kr-83m generator for KATRIN.
    Journal of Instrumentation. Roč. 13, č. 4 (2018), č. článku P04018. ISSN 1748-0221. E-ISSN 1748-0221
    R&D Projects: GA MŠMT LTT18021; GA MŠMT LM2015056
    Institutional support: RVO:61389005
    Keywords : detector alignment and calibration methods * instrument optimisation * spectrometers
    OECD category: Nuclear physics
    Impact factor: 1.366, year: 2018

    Monoenergetic conversion electrons from the Kr-83m isomeric state have been proven to be useful in the calibration of several tritium neutrino mass and dark matter experiments. In this paper the design and characteristics of the gaseous Kr-83m generator, including the Rb-83/Kr-83m source behavior in tritium, for the KATRIN experiment are presented. Using Si(Li) and silicon drift detectors (SDD) detectors, the half-life of the Kr-83m isomeric state was measured to be 1.8620 +/- 0.0019 h.
    Permanent Link: http://hdl.handle.net/11104/0283728

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.