Experimental and numerical research of mechanical behaviour of titanium alloy hip implant

  1. 1.
    0481735 - ÚGN 2018 RIV HR eng J - Journal Article
    Čolić, K. - Sedmak, A. - Legweel, K. - Milošević, M.I. - Nitrović, N. - Mišković, Ž. - Hloch, Sergej
    Experimental and numerical research of mechanical behaviour of titanium alloy hip implant.
    Technicki vjesnik - Technical Gazette. Roč. 24, č. 3 (2017), s. 709-713. ISSN 1330-3651. E-ISSN 1848-6339
    Institutional support: RVO:68145535
    Keywords : biomedical implant * digital image correlation * finite element method * Ti-6Al-4V alloy
    OECD category: Mechanical engineering
    Impact factor: 0.686, year: 2017
    https://hrcak.srce.hr/index.php?id_clanak_jezik=269849&show=clanak

    Experimental and numerical research of mechanical behaviour of hip replacement implant is presented. Experimental analysis was based on standard methods for testing mechanical properties and application of Digital Image Correlation (DIC) technique, while the Finite Element Method (FEM) was used for numerical simulation. A hip prosthesis extracted from a patient after revision surgery was used as the specimen. An adapter for the implant testing was constructed to enable the condition of a proximally loosened stem. During FEM analysis, a solid model was made of the assembly. An analysis of presented results indicated good agreement of experimental and numerical research.
    Permanent Link: http://hdl.handle.net/11104/0277242

     