Number of the records: 1  

An Improved Apparatus for Vapour-Liquid Equilibria Measurement in Polymer + Solvent Systems at Higher Temperatures: a Study of the Water + Poly(ethylene glycol) System.

  1. 1.
    0479630 - ÚCHP 2018 RIV NL eng J - Journal Article
    Pavlíček, Jan - Bogdanić, Grozdana - Wichterle, Ivan
    An Improved Apparatus for Vapour-Liquid Equilibria Measurement in Polymer + Solvent Systems at Higher Temperatures: a Study of the Water + Poly(ethylene glycol) System.
    Fluid Phase Equilibria. Roč. 454, 25 DEC (2017), s. 111-115. ISSN 0378-3812. E-ISSN 1879-0224
    R&D Projects: GA ČR GA15-19542S
    Institutional support: RVO:67985858
    Keywords : correlation * experimental data * polymer-solvent system
    OECD category: Physical chemistry
    Impact factor: 2.197, year: 2017

    Recently we designed an apparatus for total pressure measurement in binary polymer + solvent systems. Accurate as it is, the apparatus is limited by its direct capacitance pressure sensor that can operate only at 318.15 K. In this study we tested a new setup that overcomes this limitation with a pressure transducer (MKS Baratron), which can be heated up to 430 K. The main improvement is that the new setup has eliminated solvent vapour condensation in the pressure measurement unit. Other parts of the equipment can easily be kept at temperatures higher than the system temperature with heating tapes. Unlike the ebulliometric method, which is limited by problems arising from the boiling of viscous mixtures, now we can study solutions with higher polymer concentrations. We tested the new setup by measuring the total pressure of binary poly(ethylene glycol) + water mixtures at six isotherms in the (308.15–343.15) K range. Additionally we determined new data for three isotherms (343.15, 353.15, and 363.15) K using ebulliometric method. We then correlated our measurements with both methods using the UNIQUAC–free volume model and compared them with available published data for isotherms 313.15 K and 333.15 K. The agreement with our static measurement results was excellent.
    Permanent Link: http://hdl.handle.net/11104/0275604

     
    FileDownloadSizeCommentaryVersionAccess
    0479630.pdf1387.6 KBAuthor’s postprintopen-access
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.