Number of the records: 1  

Iodine Absorption Cells Purity Testing

  1. 1.
    0468639 - ÚPT 2017 RIV CH eng J - Journal Article
    Hrabina, Jan - Zucco, M. - Philippe, Ch. - Pham, Minh Tuan - Holá, Miroslava - Acef, O. - Lazar, Josef - Číp, Ondřej
    Iodine Absorption Cells Purity Testing.
    Sensors. Roč. 17, č. 1 (2017), s. 1-13, č. článku 17010102. E-ISSN 1424-8220
    R&D Projects: GA ČR GA15-18430S; GA MŠMT(CZ) LO1212; GA MŠMT ED0017/01/01
    Institutional support: RVO:68081731
    Keywords : iodine cells * absorption spectroscopy * laser spectroscopy * laser standards * frequency stability
    OECD category: Optics (including laser optics and quantum optics)
    Impact factor: 2.475, year: 2017

    This article deals with the evaluation of the chemical purity of iodine-filled absorption cells and the optical frequency references used for the frequency locking of laser standards. We summarize the recent trends and progress in absorption cell technology and we focus on methods for iodine cell purity testing. We compare two independent experimental systems based on the laser-induced fluorescence method, showing an improvement of measurement uncertainty
    by introducing a compensation system reducing unwanted influences. We show the advantages
    of this technique, which is relatively simple and does not require extensive hardware equipment. As an alternative to the traditionally used methods we propose an approach of hyperfine transitions’ spectral linewidth measurement. The key characteristic of this method is demonstrated on a set of testing iodine cells. The relationship between laser-induced fluorescence and transition linewidth methods will be presented as well as a summary of the advantages and disadvantages of the proposed technique (in comparison with traditional measurement approaches).
    Permanent Link: http://hdl.handle.net/11104/0266445

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.