Number of the records: 1  

In situ calibration of acoustic emission transducers by time reversal method

  1. 1.
    0456674 - ÚT 2017 RIV NL eng J - Journal Article
    Kober, Jan - Převorovský, Zdeněk - Chlada, Milan
    In situ calibration of acoustic emission transducers by time reversal method.
    Sensors and Actuators A - Physical. Roč. 240, April (2016), s. 50-56. ISSN 0924-4247. E-ISSN 1873-3069
    Institutional support: RVO:61388998
    Keywords : time reversed acoustics * calibration * in situ * acoustic emission
    Subject RIV: BI - Acoustics
    Impact factor: 2.499, year: 2016
    http://ac.els-cdn.com/S0924424716300334/1-s2.0-S0924424716300334-main.pdf?_tid=0acf4736-ef6d-11e5-b826-00000aacb362&acdnat=1458568911_1c21eda9762b905a684ff939463ef3fe

    Conventional methods of calibration of acoustic emission transducers require an already calibratedreference transducer and/or special test specimen. Moreover the calibration results are not directlytransferrable to practice, since they are dependent on bonding conditions and mechanical propertiesof the medium. In this paper a reference-free in situ calibration method is proposed. It is based on acombination of time reversed acoustics and reciprocal calibration and allows accurate determination oftransducers’ frequency responses. The method was developed and successfully experimentally verifiedassuming that frequency response of used transducers is the same for transmission and reception. Sub-sequently the analysis was extended considering irreversible transducers and applied to experimentaldata. Of five tested transducers, three were found reversible and two demonstrated resonance frequencyshift in lower frequency region between reception and transmission.
    Permanent Link: http://hdl.handle.net/11104/0258488

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.