Number of the records: 1  

Absolute Sensitivity Calibration of Visible Spectroscopic Diagnostic and Temporal Evolution of First Window Transmittance at the COMPASS Tokamak

  1. 1.
    0423079 - ÚFP 2014 RIV CZ eng C - Conference Paper (international conference)
    Naydenkova, Diana - Weinzettl, Vladimír - Stöckel, Jan - Háček, Pavel - Imríšek, Martin
    Absolute Sensitivity Calibration of Visible Spectroscopic Diagnostic and Temporal Evolution of First Window Transmittance at the COMPASS Tokamak.
    Proceedings of the 22nd Annual Conference of Doctoral Students – WDS 2013. Vol. 2. Prague: MATFYZPRESS, 2013 - (Šafránková, J.; Pavlů, J.), s. 97-103. WDS. ISBN 978-80-7378-251-1.
    [Annual Conference of Doctoral Students – WDS 2013 /22./. Praha (CZ), 04.06.2013-07.06.2013]
    R&D Projects: GA MŠMT(CZ) LM2011021
    Institutional support: RVO:61389021
    Keywords : COMPASS * tokamak * Visible Spectroscopic Diagnostic * Absolute Sensitivity * Calibration
    Subject RIV: BL - Plasma and Gas Discharge Physics
    http://www.mff.cuni.cz/veda/konference/wds/proc/pdf13/WDS13_215_f2_Naydenkova.pdf

    An absolute calibration of spectroscopic systems is required to measure many important plasma parameters in tokamak discharges like impurity inflow, recycling and the effective ion charge. The procedure of the absolute calibration is described and technical issues and experimental limitations connected with the tokamak operation are discussed here. Namely, an influence of the first window transmittance and its evolution over experimental campaigns on the validity of the absolute calibration is investigated