Number of the records: 1  

Correction of artifacts in diffusion-weighted MR images

  1. 1.
    0370945 - ÚPT 2012 RIV US eng C - Conference Paper (international conference)
    Marcon, P. - Bartušek, Karel - Kořínek, Radim - Dokoupil, Zdeněk
    Correction of artifacts in diffusion-weighted MR images.
    34th International Conference on Telecommunications and Signal Processing - TSP 2011. Piscataway: IEEE, 2011, s. 391-397. ISBN 978-1-4577-1410-8.
    [International Conference on Telecommunications and Signal Processing /34./ - TSP 2011. Budapest (HU), 18.08.2011-20.08.2011]
    R&D Projects: GA MŠMT ED0017/01/01; GA ČR GAP102/11/0318
    Institutional research plan: CEZ:AV0Z20650511
    Keywords : artifact * diffusion coefficient * eddy currents * gradient * magnetic resonance
    Subject RIV: JA - Electronics ; Optoelectronics, Electrical Engineering

    The authors describe the techniques for measuring diffusion coefficients and creating DWI and DTI images. In order to increase the accuracy of diffusion coefficient measurement, it is necessary to eliminate the influence of artifacts such as the influence of magnetic field inhomogeneity, eddy currents, or temperature. For these artifact sources, the article presents experimental detection of the magnitude of errors, which is realized using the three measurements method, and proposes correction possibilities. The results are compared with diffusion measurements at different sizes of 6-factors.
    Permanent Link: http://hdl.handle.net/11104/0204616

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.