Number of the records: 1  

Correction of sample tilt in FIB instruments

  1. 1.
    0358590 - ÚPT 2012 RIV NL eng J - Journal Article
    Oral, Martin - Lencová, Bohumila
    Correction of sample tilt in FIB instruments.
    Nuclear Instruments & Methods in Physics Research Section A. Roč. 645, č. 1 (2011), s. 130-135. ISSN 0168-9002. E-ISSN 1872-9576
    R&D Projects: GA AV ČR IAA100650805
    Institutional research plan: CEZ:AV0Z20650511
    Keywords : sample tilt * extraction field * correction of elliptical beam shape * calculation of intensity distribution * focused ion beams
    Subject RIV: JA - Electronics ; Optoelectronics, Electrical Engineering
    Impact factor: 1.207, year: 2011

    Many particle optical instruments use ion beams focused onto a tilted sample and the rotationally symmetrical beam creates an elliptical spot. Furthermore, an additional electrostatic field may be present near the titled sample, an extracting field of a detector or a spectrometer or a retarding field on the sample, for example. As a result, the dipole and quadrupole components of the field near the axis of the column cause an additional astigmatic focusing of the beam, which leads to even more elongated shape of the spot. A method of determining a correction has been developed, which allows to compensate for both the ellipticity due to the tilt and the additional astigmatism and to achieve a circular spot on the titled sample. The compensation is provided by a suitable setting of two correcting quadruples and by refocusing the objective lens. The required settings of the quadrupole excitations and objective lens are obtained by a paraxial optimization.
    Permanent Link: http://hdl.handle.net/11104/0196576

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.