Number of the records: 1  

Development and performance test of picosecond pulse X-ray excited streak camera system for scintillator characterization

  1. 1.
    0353670 - FZÚ 2011 RIV JP eng J - Journal Article
    Yanagida, T. - Fujimoto, Y. - Yoshikawa, A. - Yokota, Y. - Kamada, K. - Pejchal, Jan - Chani, V. - Kawaguchi, N. - Fukuda, K. - Uchiyama, K. - Mori, K. - Kitano, K. - Nikl, Martin
    Development and performance test of picosecond pulse X-ray excited streak camera system for scintillator characterization.
    Applied Physics Express. Roč. 3, č. 5 (2010), 056202/1-056202/3. ISSN 1882-0778. E-ISSN 1882-0786
    Institutional research plan: CEZ:AV0Z10100521
    Keywords : scintillation decay * streak camera * picosecond X-ray pulse * BaF2
    Subject RIV: BM - Solid Matter Physics ; Magnetism
    Impact factor: 2.747, year: 2010

    To observe time and wavelength-resolved scintillation events, picosecond pulse X-ray excited streak camera system is developed. The wavelength range spreads from VUV to near infrared region (110-900 nm) and the instrumental response function is around 80 ps. This work describes the principle of the newly developed instrument and the first performance test using BaF2 single crystal scintillator.
    Permanent Link: http://hdl.handle.net/11104/0192847

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.