Number of the records: 1
Detection of bisphenol A using a novel surface plasmon resonance biosensor
- 1.0350304 - ÚFE 2011 RIV DE eng J - Journal Article
Hegnerová, Kateřina - Piliarik, Marek - Šteinbachová, M. - Flegelová, Z. - Černohorská, H. - Homola, Jiří
Detection of bisphenol A using a novel surface plasmon resonance biosensor.
Analytical and Bioanalytical Chemistry. Roč. 398, č. 5 (2010), s. 1963-1966. ISSN 1618-2642. E-ISSN 1618-2650
R&D Projects: GA AV ČR KAN200670701; GA MŠMT OC09058
Institutional research plan: CEZ:AV0Z20670512
Keywords : surface plasmon resonance biosensor * bisphenol A * endocrine disruptor
Subject RIV: JB - Sensors, Measurment, Regulation
Impact factor: 3.841, year: 2010
In this work we present a compact surface plasmon resonance (SPR) biosensor for the detection of bisphenol A (BpA), an endocrine-disrupting chemical. The detection of BpA in PBS and wastewater samples was performed using the binding inhibition detection format. The limits of detection were determined to be 0.08 ng/ml and 0.14 ng/ml for BpA in PBS and wastewater samples, respectively. The results obtained using the SPR sensor were compared with those obtained with the ELISA and HPLC methods.
Permanent Link: http://hdl.handle.net/11104/0190339
Number of the records: 1