Number of the records: 1  

Fiber-optic pH detection in small volumes of biosamples

  1. 1.
    0349751 - ÚFE 2011 RIV DE eng J - Journal Article
    Kašík, Ivan - Mrázek, Jan - Martan, Tomáš - Pospíšilová, Marie - Podrazký, Ondřej - Matějec, Vlastimil - Hoyerová, Klára - Kamínek, Miroslav
    Fiber-optic pH detection in small volumes of biosamples.
    Analytical and Bioanalytical Chemistry. Roč. 398, č. 5 (2010), s. 1883-1889. ISSN 1618-2642. E-ISSN 1618-2650
    R&D Projects: GA MŠMT(CZ) LC06034; GA ČR(CZ) GP102/08/P639; GA AV ČR IAA600380805
    Institutional research plan: CEZ:AV0Z20670512; CEZ:AV0Z50380511
    Keywords : optical fiber sensor * plants * pH
    Subject RIV: JB - Sensors, Measurment, Regulation
    Impact factor: 3.841, year: 2010

    The paper deals with the fiber-optic detection of the pH values of droplets of plant xylem exudate based on fluorescence intensity measurements. For this purpose, novel V-taper sensing probes with a minimum diameter of around 8 μm were prepared that enable the delivery of fluorescence signal from the detection site on the taper tip to the detector. The taper tips were coated with pH-sensitive HPTS and a reference Ru-phen dichloride immobilized in a xerogel layer. Using the developed V-taper sensing probes, the pH values of in vivo and in vitro samples of small volumes (6 μl) of exudate were measured.
    Permanent Link: http://hdl.handle.net/11104/0189906

     
    FileDownloadSizeCommentaryVersionAccess
    UFE 0349751.pdf7352.3 KBOtherrequire
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.