Number of the records: 1
Coupling of hydrodynamically closed large bore capillary isotachophoresis with electrospray mass spectrometry
- 1.0343570 - ÚIACH 2011 RIV NL eng J - Journal Article
Tomáš, Roman - Koval, M. - Foret, František
Coupling of hydrodynamically closed large bore capillary isotachophoresis with electrospray mass spectrometry.
Journal of Chromatography A. Roč. 1217, č. 25 (2010), s. 4144-4149. ISSN 0021-9673. E-ISSN 1873-3778
R&D Projects: GA AV ČR IAA400310703; GA MŠMT LC06023
Institutional research plan: CEZ:AV0Z40310501
Keywords : isotachophoresis * electrophoresis * mass spectrometry
Subject RIV: CB - Analytical Chemistry, Separation
Impact factor: 4.194, year: 2010
In this work we have modified the commercial coupled column isotachophoresis system for direct connection to an ion trap mass spectrometer. The design included attachment of an elution block with a short capillary transfer line directing the separated zones towards the mass spectrometer. The modification further included separation of the injection and electrode blocks from the separation columns by semipermeable membranes eliminating unwanted fluid movements in the wide bore fluoropolymer separation capillaries.
Permanent Link: http://hdl.handle.net/11104/0186020