Influence of the sample mounting on thermal conductance measurements using PPMS TTO option

  1. 1.
    0329997 - FZÚ 2010 RIV GB eng J - Journal Article
    Šebek, Josef - Šantavá, Eva
    Influence of the sample mounting on thermal conductance measurements using PPMS TTO option.
    [Vliv uspořádání vzorku na měření tepelné vodivosti při použití PPMS TTO.]
    Journal of Physics: Conference Series. Roč. 150, - (2009), 012044/1-012044/4. ISSN 1742-6588. E-ISSN 1742-6596
    Institutional research plan: CEZ:AV0Z10100520
    Keywords : thermal transport
    Subject RIV: BM - Solid Matter Physics ; Magnetism

    The TTO implemented into the PPMS measures the thermal transport properties of materials. Recently, discrepancies up to 30% in measured quantities at 390 K have been reported. We critically analyze the experimental method used to measure and show possible sources of problems.

    Tepelný přenos v materiálech se v PPMS měří s využitím nadstavby TTO. Nedávno byly publikovány až 30% nesrovnalosti v těchto měřeních při 390 K. V příspěvku kriticky analyzujeme použitou experimentální metodu a ukazujeme na možné příčiny.