System zones in capillary zone electrophoresis: Moving boundaries caused by freely migrating hydroxide ions

  1. 1.
    0021944 - ÚIACH 2006 RIV DE eng J - Journal Article
    Beckers, J. L. - Urbánek, Marek - Boček, Petr
    System zones in capillary zone electrophoresis: Moving boundaries caused by freely migrating hydroxide ions.
    [Systémové zóny v kapilární zónové elektroforéze. Pohyblivá rozhraní způsobená volně migrujícími hydroxidovými ionty.]
    Electrophoresis. Roč. 26, č. 10 (2005), s. 1869-1873. ISSN 0173-0835. E-ISSN 1522-2683
    R&D Projects: GA AV ČR IAA4031103
    Institutional research plan: CEZ:AV0Z40310501
    Keywords : background electrolyte * capillary electrophoresis * system zones
    Subject RIV: CB - Analytical Chemistry, Separation
    Impact factor: 3.850, year: 2005

    We present theoretical and experimental data indicating that anionic system zones (SZs), due to free migrating hydroxide anions, can be expected in background electrolytes (BGEs) with a low buffer capacity. In the system containing completely unbuffered BGEs the hydroxide ions migrate freely through the system and cause stepwise disturbances in the baseline of the detector trace. This type of SZs do not contribute to the electromigation dispersion of the zones of the analytes.

    Jsou prezentována teoretická a experimentální data ukazující, že vznik anionických systémových zón vzniklých v důsledku volně migrujících hydroxidových iontů lze očekávat v základních elektrolytech s nízkou pufrační kapacitou. V systému se zcela nepufrovaným základním elektrolytem migrují hydroxidové ionty systémem zcela volně a narušují základní linii detekčního záznamu. Tento typ systémových zón nepřispívá k elektromigrační disperzi zón analytů.