Number of the records: 1  

Construction and Application of Flow Enzymatic Biosensor Based of Silver Solid Amalgam Electrode for Determination of Sarcosine

  1. 1.
    0446079 - ÚFCH JH 2016 RIV DE eng J - Journal Article
    Josypčuk, Oksana - Barek, J. - Josypčuk, Bohdan
    Construction and Application of Flow Enzymatic Biosensor Based of Silver Solid Amalgam Electrode for Determination of Sarcosine.
    Electroanalysis. Roč. 27, č. 11 (2015), s. 2559-2566. ISSN 1040-0397. E-ISSN 1521-4109
    R&D Projects: GA ČR GBP206/12/G151; GA ČR GAP206/11/1638
    Institutional support: RVO:61388955
    Keywords : biosensors * sarcosine * silver solid amalgam electrode
    Subject RIV: CG - Electrochemistry
    Impact factor: 2.471, year: 2015

    In this paper, the flow amperometric enzymatic biosensor based on polished silver solid amalgam electrode for determination of sarcosine in model sample under flow injection analysis conditions is presented. The biosensor works on principle of electrochemical detection of oxygen decrease during enzymatic reaction which is directly proportional to the concentration of sarcosine in sample. The whole preparation process takes about 3 h. The RSD of repeatability of 10 consecutive measurements is 1.6% (csarcosine=1.0x10-4 moldm-3). Under optimal conditions the calibration dependence was linear in the range 7.5x10-6 – 5.0x10-4 moldm-3 and limit of detection was 2.0x10-6 moldm-3.
    Permanent Link: http://hdl.handle.net/11104/0248062

     
     