Number of the records: 1  

SiC and Si detectors comparison for high carbon energy spectrometry

  1. 1.
    0587877 - ÚJF 2025 RIV GB eng J - Journal Article
    Torrisi, L. - Havránek, Vladimír - Macková, Anna - Torrisi, A. - Cutroneo, Mariapompea
    SiC and Si detectors comparison for high carbon energy spectrometry.
    Journal of Instrumentation. Roč. 19, č. 7 (2024), č. článku P07005. ISSN 1748-0221. E-ISSN 1748-0221
    R&D Projects: GA MŠMT EF16_013/0001812
    Institutional support: RVO:61389005
    Keywords : Interaction of radiation with matter * Radiation-hard detectors * Solid state detectors
    OECD category: Nuclear physics
    Impact factor: 1.3, year: 2022
    Method of publishing: Open access
    https://doi.org/10.1088/1748-0221/19/07/P07005

    An innovative SiC Schottky junction and a traditional p-n Si surface barrier detector have been compared to detect carbon ions with MeVs kinetic energy. To this, a comparison was performed during Rutherford backscattering spectrometry (RBS) using 2-10 MeV carbon ion beams. The energy resolution and detection efficiency for RBS analysis using the two detectors and their detection electronics are presented. The detector parameters dependencies on the surface passivating layers, ion energy and current dependence, ion penetration depth, detection efficiency, energy resolution, and others are discussed. The comparison of RBS analysis with SiC and Si is investigated highlighting the advantages and disadvantages of using SiC with respect to the traditional Si surface barrier detectors. The two detectors employed for proton, helium and carbon RBS spectrometry of different targets have been also compared on the base of the literature data.
    Permanent Link: https://hdl.handle.net/11104/0354916

     
    FileDownloadSizeCommentaryVersionAccess
    0587877.pdf03 MBCC licencePublisher’s postprintopen-access
     