Number of the records: 1
Atomization of As and Se volatile species in a dielectric barrier discharge atomizer after hydride generation: Fate of analyte studied by selected ion flow tube mass spectrometry
- 1.0547833 - ÚIACH 2023 RIV NL eng J - Journal Article
Kratzer, Jan - Lacko, Michal - Dryahina, Kseniya - Matoušek, Tomáš - Španěl, Patrik - Dědina, Jiří
Atomization of As and Se volatile species in a dielectric barrier discharge atomizer after hydride generation: Fate of analyte studied by selected ion flow tube mass spectrometry.
Analytica Chimica Acta. Roč. 1190, JAN (2022), č. článku 339256. ISSN 0003-2670. E-ISSN 1873-4324
R&D Projects: GA ČR(CZ) GA21-05285S; GA ČR(CZ) GA21-25486S
Institutional support: RVO:68081715 ; RVO:61388955
Keywords : dielectric barrier discharge atomizer * hydride atomization mechanism * selected ion flow tube mass spectrometry
OECD category: Analytical chemistry; Physical chemistry (UFCH-W)
Impact factor: 6.2, year: 2022
Method of publishing: Limited access
Atomization of As and Se hydrides and their methylated analogues in a dielectric barrier discharge (DBD) plasma atomizer was investigated. Selected ion flow tube mass spectrometry (SIFT-MS) was chosen as a detector being capable of selective detection of non-atomized original volatile species allowing thus direct quantification of atomization efficiency. Apart from investigation of analyte atomization, the SIFT-MS detector is capable of quantitative determination of water vapor content being either transported to, or produced in the atomizer.
Permanent Link: http://hdl.handle.net/11104/0323993
File Download Size Commentary Version Access 0547833.pdf 3 1.1 MB Author´s preprint require
Number of the records: 1