Number of the records: 1
On UV-Vis Spectroelectrochemistry of Luteolin-7-O-Glucoside
- 1.0505245 - ÚFCH JH 2021 RIV CZ eng C - Conference Paper (international conference)
Castano, A. - Sokolová, Romana - Degano, I.
On UV-Vis Spectroelectrochemistry of Luteolin-7-O-Glucoside.
XXXIX. Moderní elektrochemické metody. Sborník přednášek. Ústí nad Labem: Best Servis, 2019 - (Navrátil, T.; Fojta, M.; Schwarzová, K.), s. 31-35. ISBN 978-80-905221-7-6.
[Moderní elektrochemické metody /39./. Jetřichovice (CZ), 20.05.2019-24.05.2019]
R&D Projects: GA ČR(CZ) GA19-03160S
Institutional support: RVO:61388955
Keywords : spectroelectrochemistry * oxidation * luteolin * antioxidants
OECD category: Physical chemistry
The UV-Vis spectroelectrochemical cell containing an optically transparent platinum net used as the working electrode was tested in two regimes of electrochemical measurement, cyclic voltammetry and chronoamperometry, respectively. A three electrode setup used silver quasi reference electrode and platinum electrode serving as a counter electrode. The calibration of the spectroelectrochemical cell was performed for one-electron oxidation processes as well as for two-electron oxidations. The spectral changes during the oxidation of luteolin-7-O-Glucoside gave the information about the electroactive chromophores in molecule. Resulted spectra were compared with luteolin. The identification of products of oxidation was done by complementary analytical techniques HPLC-DAD and HPLC-ESI-MS/MS.
Permanent Link: http://hdl.handle.net/11104/0296729
File Download Size Commentary Version Access 0505245.pdf 1 290.8 KB Author’s postprint open-access