Number of the records: 1
Resampling in magnetic resonance spectroscopy-A less model-dependent quantitation quality assessment method
- 1.0477997 - ÚPT 2018 RIV SK eng C - Conference Paper (international conference)
Jabłoński, Michal - Starčuková, Jana - Starčuk jr., Zenon
Resampling in magnetic resonance spectroscopy-A less model-dependent quantitation quality assessment method.
11th International Conference on Measurement, Measurement 2017. Proceedings. Bratislava: Institute of Measurement Science SAS, 2017, s. 193-196. ISBN 978-80-972629-0-7.
[Measurement 2017. International Conference on Measurement /11./. Smolenice (SK), 29.05.2017-31.05.2017]
R&D Projects: GA MŠMT(CZ) LO1212; GA MŠMT ED0017/01/01
EU Projects: European Commission(XE) 316679
Institutional support: RVO:68081731
Keywords : magnetic resonance spektroscopy * quantitation * fitting Quality Assessment
OECD category: Medical engineering
In this paper we propose a novel method for magnetic resonance spectroscopy (MRS) fitting quality assessment based on time domain resampling of acquired spectra and bootstrapping of the obtained results. The authors introduce two resampling strategies and example results are shown. The proposed method requires MRS data without averaging.