Number of the records: 1
Quantum dots as chemiluminescence enhancers tested by sequential injection technique: Comparison of flow and flow-batch conditions
- 1.0468753 - ÚIACH 2018 RIV NL eng J - Journal Article
Sklenářová, H. - Voráčová, Ivona - Chocholouš, P. - Polášek, M.
Quantum dots as chemiluminescence enhancers tested by sequential injection technique: Comparison of flow and flow-batch conditions.
Journal of Luminescence. Roč. 184, APR (2017), s. 235-241. ISSN 0022-2313. E-ISSN 1872-7883
Institutional support: RVO:68081715
Keywords : quantum dots * chemiluminescence * sequentialinjectionanalysis
OECD category: Analytical chemistry
Impact factor: 2.732, year: 2017
The effect of Quantum Dots (QDs) with different emission wavelengths
(520–640nm) and different surface modifications (mercaptopropionic, mercaptoundecanoic, thioglycolic acids and
mercaptoethylamine) on permanganate-induced and luminol–hydrogen peroxide chemiluminescence (CL) was studied in detail by a sequential injection technique using a spiral detection flow cell and a flow- batch detection cell operated in flow and stop-flow modes.
Permanent Link: http://hdl.handle.net/11104/0266579
File Download Size Commentary Version Access 0468753_IR.pdf 3 471.2 KB Other require
Number of the records: 1