Number of the records: 1  

The Use of the Silver Solid Amalgam Electrode for Voltammetric Determination of 9-Nitroanthracene

  1. 1.
    0446745 - ÚFCH JH 2017 RIV US eng J - Journal Article
    Navrátil, Tomáš - Nováková, Kateřina - Barek, J. - Vyskočil, V. - Chýlková, J.
    The Use of the Silver Solid Amalgam Electrode for Voltammetric Determination of 9-Nitroanthracene.
    Analytical Letters. Roč. 49, č. 1 (2016), s. 37-48. ISSN 0003-2719. E-ISSN 1532-236X
    R&D Projects: GA ČR(CZ) GAP208/12/1645
    Institutional support: RVO:61388955
    Keywords : 9-Nitroanthracene * Silver solid amalgam electrode * Voltammetry
    Subject RIV: CG - Electrochemistry
    Impact factor: 1.150, year: 2016

    The voltammetric behavior of 9-nitroanthracene was investigated at a mercury meniscus modified silver solid amalgam electrode (m-AgSAE; inner diameter of 0.5 millimeter) by differential pulse voltammetry (DPV) and adsorptive stripping DPV (DPAdSV). Optimum conditions were found for the determination of this analyte in Britton-Robinson buffer (pH = 10.5): pulse width of 80 milliseconds, pulse height of −50 millivolts, initial potential (Ein) = −100 millivolts, final potential (Efin) = −1200, potential of accumulation (Eacc) = −100 millivolts, scan rate of 20 millivolts per second. The limit of detection was found to be 0.7 nanomole per liter (tacc = 60 seconds) by DPAdSV at m-AgSAE. The method was successfully applied for the determination of 9-nitroanthracene in drinking and river water.
    Permanent Link: http://hdl.handle.net/11104/0248719

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.