Active angular alignment of gauge block in system for contactless gauge block calibration

  1. 1.
    0434167 - ÚPT 2015 RIV US eng C - Conference Paper (international conference)
    Buchta, Zdeněk - Šarbort, Martin - Řeřucha, Šimon - Hucl, Václav - Čížek, Martin - Lazar, Josef - Číp, Ondřej
    Active angular alignment of gauge block in system for contactless gauge block calibration.
    Optical Micro- and Nanometrology V. (Proceedings of SPIE 9132). Bellingham: SPIE, 2014, 91321J:1-7. ISSN 0277-786X.
    [Optical Micro- and Nanometrology /5./. Brussels (BE), 15.04.2014-17.04.2014]
    R&D Projects: GA TA ČR(CZ) TA03010663; GA ČR GP102/09/P293; GA ČR GAP102/10/1813; GA MŠMT ED0017/01/01; GA MŠMT(CZ) LC06007; GA MŠMT(CZ) LO1212
    Institutional support: RVO:68081731
    Keywords : low-coherence interferometry * gauge block * metrology
    Subject RIV: BH - Optics, Masers, Lasers

    This paper presents a method for active angular alignment of gauge block implemented in a system for automatic contactless calibration of gauge blocks designed at ISI ASCR. The system combines low-coherence interferometry and laser interferometry, where the first identifies the gauge block sides position and the second one measures the gauge block length itself. A crucial part of the system is the algorithm for gauge block alignment to the measuring beam which is able to compensate the gauge block lateral and longitudinal tilt up to 0.141 mrad. The algorithm is also important for the gauge block position monitoring during its length measurement.