Number of the records: 1  

Electron line shape of the KATRIN monitor spectrometer

  1. 1.
    0425299 - ÚJF 2014 RIV GB eng J - Journal Article
    Slezák, Martin - Bauer, S. - Dragoun, Otokar - Erhard, M. - Schlosser, K. - Špalek, Antonín - Vénos, Drahoslav - Zbořil, M.
    Electron line shape of the KATRIN monitor spectrometer.
    Journal of Instrumentation. Roč. 8, DEC (2013), T12002. ISSN 1748-0221. E-ISSN 1748-0221
    R&D Projects: GA ČR(CZ) GAP203/12/1896
    Institutional support: RVO:61389005
    Keywords : spectrometers * analysis and statistical methods * detector alignment and calibration method * KATRIN
    Subject RIV: BG - Nuclear, Atomic and Molecular Physics, Colliders
    Impact factor: 1.526, year: 2013

    Conversion electrons emitted from Kr-83m implanted into a solid substrate will serve as a powerful tool for monitoring of the energy scale stability in the KATRIN neutrino experiment. An appropriate description of the conversion line shape is essential to determine the energy of the emitted electrons. It is shown that the Doniach-.Sunjic line shape gives a significantly better fit to the conversion electron spectra than the previously used double Voigt model. The electron spectra were obtained with the KATRIN MAC-E filter monitor spectrometer.
    Permanent Link: http://hdl.handle.net/11104/0231177

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.