Number of the records: 1  

Voltammetric monitoring of electrochemical reduction of riboflavin using silver solid amalgam electrodes

  1. 1.
    0377573 - ÚFCH JH 2013 RIV GB eng J - Journal Article
    Bandžuchová, L. - Šelešovská, R. - Navrátil, Tomáš - Chýlková, J. - Novotný, L.
    Voltammetric monitoring of electrochemical reduction of riboflavin using silver solid amalgam electrodes.
    Electrochimica acta. Roč. 75, July 2012 (2012), s. 316-324. ISSN 0013-4686. E-ISSN 1873-3859
    R&D Projects: GA AV ČR IAA400400806; GA ČR GAP208/12/1645; GA ČR GAP206/11/1638
    Institutional support: RVO:61388955
    Keywords : riboflavin * determination * amalgam electrode
    Subject RIV: CG - Electrochemistry
    Impact factor: 3.777, year: 2012

    The voltammetric behavior of riboflavin (RF), an important compound belonging to the group of vitamins B, has been investigated using mercury meniscus modified electrode (m-AgSAE) and polished silver solid amalgam electrode (p-AgSAE). Obtained results were compared with those obtained using hanging mercury drop electrode (HMDE). Cyclic voltammetry (CV), direct current voltammetry (DCV) and differential pulse voltammetry (DPV) have been applied for investigation of electrochemical behavior of RF on the above mentioned electrodes. Optimum conditions for DPV determination of RF have been found in acetate buffer of pH 5 for m-AgSAE and HMDE, and in Britton–Robinson (BR) buffer of pH 3 for p-AgSAE. The limits of detection amounted to 8.2 × 10−10 mol L−1 (m-AgSAE) and 1.3 × 10−9 mol L−1 (p-AgSAE). DPV with optimized parameters was utilized for successful determination of RF in 2 vitamin preparations.
    Permanent Link: http://hdl.handle.net/11104/0209694

     
     