Number of the records: 1  

Detection of bisphenol A using a novel surface plasmon resonance biosensor

  1. 1.
    0356064 - ÚFE 2011 CZ eng A - Abstract
    Hegnerová, Kateřina - Piliarik, Marek - Homola, Jiří
    Detection of bisphenol A using a novel surface plasmon resonance biosensor.
    EUROPT(R)ODE X. Vol. XConference on Optical Chemical Sensors and Biosensors. Praha: Institute of Photonics and Electronics AS CR, v.v.i, 2010 - (Homola, J.). s. 225-225. ISBN 978-80-86269-20-7.
    [EUROPT(R)ODE X – X.Conference on Optical Chemical Sensors and Biosensors. 28.03.2010-31.3.2010, Praha]
    R&D Projects: GA AV ČR KAN200670701; GA MŠMT OC09058
    Institutional research plan: CEZ:AV0Z20670512
    Keywords : surface plasmon resonance biosensor * bisphenol A * endocrine disruptor
    Subject RIV: JB - Sensors, Measurment, Regulation

    In this paper, we present a compact 6-channel surface plasmon resonance (SPR) biosensor for detection of BpA. To detect BpA, the binding inhibition detection format was used. Detection of known concentrations of BpA was performed to establish a calibration curve in buffer. Then the sensor response to BpA-spiked waste water sample was determined with SPR method and the results were compared with ELISA results. Detection limit was determined to be 0.2 ng/mL (better by a factor of 50 than ELISA).
    Permanent Link: http://hdl.handle.net/11104/0194684

     
     
Number of the records: 1  