Number of the records: 1  

Glucose determination using a re-usable enzyme-modified ion track membrane sensor

  1. 1.
    0338731 - ÚJF 2010 RIV GB eng J - Journal Article
    Fink, Dietmar - Klinkovich, I. - Bukelman, O. - Marks, R.S. - Kiv, A. - Fuks, D. - Fahrner, W. R. - Alfonta, L.
    Glucose determination using a re-usable enzyme-modified ion track membrane sensor.
    Biosensors and Bioelectronics. Roč. 24, č. 8 (2009), s. 2702-2706. ISSN 0956-5663. E-ISSN 1873-4235
    Institutional research plan: CEZ:AV0Z10480505
    Keywords : Glucose sensor * etched tracks * Ion track membranes
    Subject RIV: BG - Nuclear, Atomic and Molecular Physics, Colliders
    Impact factor: 5.429, year: 2009

    A novel concept for a glucose biosensor based on the enzyme glucose oxidase covalently linked to nanopores of etched nuclear track membranes is presented. This device can be used to detect physiologically relevant glucose concentrations between 10 mu M and 1 M. The sensitive catalytic sensor can be made re-usable due to the production of diffusible products from the oxidative biomolecular recognition event. We have demonstrated both the simplicity with which this sensor can be produced and the stability of the enzyme embedded in the nanopores. We strongly believe that the approach could be expanded for different enzymes and variable configurations of sensing and catalysis using membranes with nanopores for biocatalysis and enrichment of substrates and products. This is the first time that such devices are being used for biocatalytic transformations.
    Permanent Link: http://hdl.handle.net/11104/0182428

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.