Number of the records: 1  

Sensor for measurement of wear in total hip arthroplasty

  1. 1.
    0312562 - FZÚ 2009 RIV DE eng C - Conference Paper (international conference)
    Mandát, Dušan - Hrabovský, Miroslav - Havránek, Vítězslav - Pochmon, Michal - Rössler, T. - Gallo, J.
    Sensor for measurement of wear in total hip arthroplasty.
    [Sensor pro měření opotřebení kloubních implantátů.]
    Proceeding of the NBC - 14th Nordic-Baltic Conference on Biomedical Engineering and Medical Physics. Berlin: Springer, 2008 - (Katashev, A.; Dekhtyar, Y.; Spigulis, J.), s. 380-382. IFMBE Proceedings, vol. 20. ISBN 978-3-540-69366-6. ISSN 1680-0737.
    [Nordic-Baltic Conference on Biomedical Engineering and Medical Physics /14./. Riga (LT), 16.06.2008-20.06.2008]
    R&D Projects: GA MŠMT(CZ) 1M06002; GA MŠMT(CZ) OC 168
    Institutional research plan: CEZ:AV0Z10100522
    Keywords : cotyle ABG I * 3D topography * LabView * Matlab * total deformation
    Subject RIV: BH - Optics, Masers, Lasers

    This article treats of measurements of the total deformation of polyethylen cotyles ABG I of coxa by means of the optical topography method. We use 3D scanning topography to measure surface geometry of cotyle. The sensor consists of optical source, digital camera, rotating stage and linear stage. We use LabView to control the measurement and Matlab to compute the result.

    Článek pojednává o měření deformací polyethylenových kyčelních jamek ABG I optickými topografickými metodami. Pro měření byla použita metoda 3D skenovací topografie. Sensor je tvořen digitální kamerou, světelným zdrojem, rotačním stolkem a lineárním posuvem. Systém je řízen prostředím LabView a výpočty jsou realizovány v programu Matlab.
    Permanent Link: http://hdl.handle.net/11104/0163592

     
     
Number of the records: 1  

  This site uses cookies to make them easier to browse. Learn more about how we use cookies.