    It is known that the E-T scintillation PMT system [1] represents the most efficient detector of signal electrons in the SEM. Today, practically all types of SEMs are standardly equipped with this detector for the detection of secondary electrons (SEs). For backscattered electrons (BSEs) not only scintillation detector but also semiconductor and channel plate detectors are used. The highest detection quantum efficiency (DQE) is achieved with the scintillation detector on the basis of the yttrium aluminium garnet single crystal scintillator (YAG)[2].