    Spurný, František - Bottollier-Depois, J. F. - Vlček, Bohumil
    Spectrometry of the let LET with track etched detectors-correlation with proportional counter measured spectra.
    Radiation Measurements. Roč. 32, 1-6 (2001), s. 193-197. ISSN 1350-4487. E-ISSN 1879-0925
    R&D Projects: GA ČR GA202/01/0710; GA AV ČR KSK1010104
    Institutional research plan: CEZ:AV0Z1048901
    Keywords : experimental microdosimetry * track etched LET spectrometry * tissue equivalent proportional counter
    Subject RIV: BG - Nuclear, Atomic and Molecular Physics, Colliders
    Impact factor: 1.003, year: 2001

    The work describes the results of comparison of two experimental microdosimetry methods: tissue equivalent proportional counter, and LET spectrometer based on the chemically etched PADC track detector. The methods were compared in the region of common sensibility between 10 and 700 keV/m. It is shown that the results obtained by both methods are similar, some differences are, nevertheless, observed.