Počet záznamů: 1
Experimental and numerical research of mechanical behaviour of titanium alloy hip implant
- 1.
SYSNO ASEP 0481735 Druh ASEP J - Článek v odborném periodiku Zařazení RIV J - Článek v odborném periodiku Poddruh J Článek ve WOS Název Experimental and numerical research of mechanical behaviour of titanium alloy hip implant Tvůrce(i) Čolić, K. (RS)
Sedmak, A. (RS)
Legweel, K. (RS)
Milošević, M.I. (RS)
Nitrović, N. (RS)
Mišković, Ž. (RS)
Hloch, Sergej (UGN-S) RID, SAI, ORCIDCelkový počet autorů 7 Zdroj.dok. Technicki vjesnik - Technical Gazette , Scientific-professional journal of technical faculties of the University of Osijek. - : Sveuciliste Josipa Jurja Strossmayera u Osijeku - ISSN 1330-3651
Roč. 24, č. 3 (2017), s. 709-713Poč.str. 5 s. Forma vydání Online - E Jazyk dok. eng - angličtina Země vyd. HR - Chorvatsko Klíč. slova biomedical implant ; digital image correlation ; finite element method ; Ti-6Al-4V alloy Vědní obor RIV JQ - Strojní zařízení a nástroje Obor OECD Mechanical engineering Institucionální podpora UGN-S - RVO:68145535 UT WOS 000403788500009 DOI 10.17559/TV-20160219132016 Anotace Experimental and numerical research of mechanical behaviour of hip replacement implant is presented. Experimental analysis was based on standard methods for testing mechanical properties and application of Digital Image Correlation (DIC) technique, while the Finite Element Method (FEM) was used for numerical simulation. A hip prosthesis extracted from a patient after revision surgery was used as the specimen. An adapter for the implant testing was constructed to enable the condition of a proximally loosened stem. During FEM analysis, a solid model was made of the assembly. An analysis of presented results indicated good agreement of experimental and numerical research. Pracoviště Ústav geoniky Kontakt Lucie Gurková, lucie.gurkova@ugn.cas.cz, Tel.: 596 979 354 Rok sběru 2018 Elektronická adresa https://hrcak.srce.hr/index.php?id_clanak_jezik=269849&show=clanak
Počet záznamů: 1